Buyer's guide

Autoclave Buying Guide for Clinics in Kenya: Class N, S and B Explained

An autoclave is one of the bigger equipment purchases a clinic makes, and the wrong class is hard to fix later. Here is how to choose, specify and request quotations for a steam sterilizer.

Key takeaways

  • Small steam sterilizers fall into Class N, S and B under EN 13060, and the class decides which loads you can process. [2]
  • Choose Class B if you will sterilize wrapped, hollow or porous items, and Class N only for unwrapped solid instruments. [2]
  • Hard water, power and space affect performance, so check the manufacturer's needs before you buy. [1]
  • Plan validation, indicators and servicing from day one, because WHO recommends validation at least once a year. [1]

What is an autoclave and what does it do?

An autoclave is a steam sterilizer. It uses steam under high pressure at 121°C or above to sterilize reusable equipment so it can be used again. [1] It is for reusable instruments only. Single use items such as syringes, needles and gloves are used once and disposed of safely, and WHO warns about the risks of reusing them. [1]

An autoclave cannot rescue a dirty instrument. Cleaning is the first step, because removing soil lets steam reach the surface. [1] Budget for cleaning supplies, brushes and drying space alongside the machine.

Class N, S or B: which autoclave do you need?

EN 13060 sorts small steam sterilizers into three classes by how they remove air and what loads they handle. [2] The standard was republished in August 2025, so ask which edition a unit complies with. [3]

ClassAir removalSuitable loadsTypical fit
NGravity displacementUnwrapped, solid, non porous instruments [2]Basic needs and tighter budgets
SPartial vacuum or steam pulsingOnly the loads the manufacturer declares [2]A fixed, known range of instruments
BFull vacuum before steamSolid, hollow, porous, wrapped and unwrapped items [2]Busy clinics that use pouches or hollow items

The air removal method is built into the machine, so choose the class you will need over the next few years rather than the cheapest one today. If you consider Class S, ask for the written list of loads it is declared to handle.

WHO also describes pre vacuum, steam flush pressure pulse and gravity sterilizers, and notes that pre vacuum units need a daily Bowie-Dick test for steam penetration. [1]

Size, capacity and backup

Tabletop units suit small clinics, while larger horizontal units suit hospitals. [2] Count the instrument sets you reprocess in a busy day, allow for the cycle and cooling time, then pick a chamber that covers that load without overfilling.

Plan for failure. WHO advises having enough machines for the workload, including back up, and an alternative method if a machine stops. [1] A single autoclave with no plan for downtime puts the whole clinic at risk.

Water, power and space

  • Water: hard water can block valves and damage instruments, and some manufacturers recommend reverse osmosis, so check this before purchase, especially on a tight budget. [1]
  • Power: ask for the voltage, current and plug type, and whether a dedicated circuit or stabiliser is needed. Kenya's mains supply is 240 V at 50 Hz, so confirm the unit's rating.
  • Space: keep the minimum distances from walls that the manufacturer gives, so the unit can be ventilated and serviced. [1]

Indicators, validation and servicing

A sterilizer is only as reliable as its monitoring. WHO describes chemical indicators, biological indicators and physical checks for every process. [1] It recommends validation before first use and at least once a year, repeated after repairs. [1] For critical machines it suggests a limited technical check every 6 months and an extensive check each year. [1] WHO also notes that preventive maintenance is often skipped until the machine fails. [1]

So ask every supplier about installation and commissioning, operator training, warranty, spare parts and a service plan, and budget for indicator strips and pouches. Our supplier comparison guide shows how to weigh these offers, and our storage guide covers keeping sterilized items sound.

The Kenyan context

Kenya's Ministry of Health has published a National Infection Prevention and Control Policy for Health Care Services, now in its third edition (2021). [4] Facility checklists also differ by level, so read the current KMPDC checklist for yours, as explained in our clinic equipment checklist. A document citing the 2019 Kenya Health Facility Assessment reports that only 12% of facilities had every item for standard precautions. [5] Reliable equipment and consumables are part of closing that gap.

What to put in your quotation request

  1. The class you need (N, S or B) and the standard it must meet.
  2. Chamber volume and the loads you will process.
  3. Power supply and the water type you can provide.
  4. Accessories such as trays and baskets, plus indicator strips and pouches.
  5. Warranty, installation, operator training and service support.
  6. Your delivery location.

Example request: "We are a dental clinic in Kisumu. Please quote for one Class B tabletop autoclave, 240 V, with trays, installation and operator training, plus a year of indicator strips and pouches. Please include the warranty and service terms."

Pricing depends on class, size and service terms, so we give it in a quotation. See our products page for examples, then send your list through the contact page using our quotation guide. Related equipment is covered in our diagnostic equipment guide and waste supplies guide.

Related reading: planned maintenance for clinic equipment.

Frequently asked questions

Class N autoclaves remove air by gravity and suit unwrapped, solid instruments only. Class B autoclaves use a full vacuum before steam enters, so they can process wrapped, hollow and porous loads as well as solid items.

WHO guidance suggests a limited technical check of critical machines every 6 months and an extensive check followed by validation once a year. Follow your manufacturer's service schedule and repeat validation after repairs.

Use the water the manufacturer specifies. Hard water leaves deposits that can block valves and damage instruments, and many tabletop units call for distilled or demineralised water. Confirm this for your model before buying.

An autoclave is for reusable instruments. Single use items such as syringes, needles and gloves are used once and disposed of safely, so they are not sterilized again.

State the class, chamber size, power supply, water type, accessories, warranty, installation, operator training and service support, plus the delivery location. We usually reply within 5 to 60 minutes during business hours.

Note: this guide supports purchasing decisions. It is not clinical or infection control advice. Cloaked Innovations Limited is a non-pharmaceutical and equipment supplies partner in Kenya and does not supply medicines. Follow your facility's infection prevention and control policy and current Ministry of Health guidance.

Visit our homepage or read about us to learn more about how we support institutions across Kenya.

Sources

  1. World Health Organization and Pan American Health Organization. Decontamination and reprocessing of medical devices for health-care facilities (2016). https://www.who.int/publications/i/item/9789241549851
  2. Tuttnauer. Types of autoclave sterilizers (EN 13060 classes N, S and B). https://tuttnauer.com/knowledge-center/sterile-processing/autoclave/types-of-autoclave-sterilizers
  3. Standards.ie. EN 13060:2025 Sterilizers for medical purposes, small steam sterilizers, published 27 August 2025. https://shop.standards.ie/en-ie/standards/en-13060-2025-344383_saig_cen_cen_3653165
  4. Ministry of Health, Kenya. National Infection Prevention and Control Policy for Health Care Services, Third Edition (2021). https://www.medbox.org/document/kenya-national-infection-prevention-and-control-policy-for-health-care-services-third-edition
  5. Kenya WASH in health care facilities technical working group. Terms of reference, citing the 2019 Kenya Health Facility Assessment. https://washinhcf.org/wp-content/uploads/2025/04/Kenya_TOR-for-WASH-TWG-Editted.pdf

Ready to request a quote?

Send us your list with quantities and delivery location. We usually reply within 5 to 60 minutes during business hours.

Request a Quote Quotation guide
Chat with Cloaked Innovations

Tell us a bit about you to start the chat.